16 17OILEVELSTARTSTOPREMOVE THIS LABELSERIES STELIZERFOR USE WITH 5000LEAVE THIS LABEL ONSERIES STERILIZERFOR USE WITH 2000DAILY OPERATIONIf the quick
18 19Cassette Seal SCS001Compressor Filter SCF002Biological Filter SCF032PART RPI PART # STERILIZER PM TIPRefer to the “Installation and Monthly Maint
20 21Cassette Seal SCS029Air Compressor Filter SCF031Biological Filter SCF032PART RPI PART # STERILIZER PM TIPRefer to the “Installation and Monthly M
22 23STEP #1 - The cassette must be opened and cooled before removing the cassette seal. See Figure A.STEP #2 - Start with any of the (4) corners and
24 255.0 – SERVICE TIPS1. CASSETTE DIAGNOSTICSIf water is leaking at the drain from the cassette bay (un-derside of autoclave), then do the following:
6.0 – LCD ERROR CODES (Note: Error Codes that are not currently used by Sci-Can are listed as "Not Currently In Use" or not listedin this bo
28 296.0 – LCD ERROR CODES (continued)STAGE #3 – PRESSURIZINGThe pressure increases as the cassette temperature rises to the factory set temperatures
30 316.0 – LCD ERROR CODES (continued)STAGE #4 – STERILIZATION(continued)LCD MESSAGE CHECK CASSETTE #7 (CYCLE FAULT #7)The steam generator/boiler or
32 336.0 – LCD ERROR CODES (continued)STAGE #5 – VENTING(continued)LCD MESSAGE CYCLE FAULT #16Steam generator/boiler temperature "ran away"
34 357.0 – LCD MESSAGESAPPLIES TO ANY STAGELCD MESSAGEGARBLED OR NO READ-OUT ON THE LCDTHINGS TO CHECK• Make sure that the cable from the LCD board (R
1FIGURE 1 – STATIM 2000 & 5000 KEYPAD SYMBOLS CHARTRubber Model 2000 Unwrapped Wrapped & Plastics Air Dry Start StopRubber Heavy DutyModel 500
36 378.0 – TROUBLESHOOTING TIPSCONDITION:WATER QUALITY SENSOR TROUBLESHOOTINGApplies to the STATIM models with water quality sensoronly.THINGS TO CHEC
FIGURE 9.2 - CLEANING THE STEAM GENERATOR/BOILERFIGURE 9.1 - STEAM GENERATOR/BOILER38 399.0 – REFERENCE BLOCKS • To check the thermal fuse continuity
FIGURE 9.3 - STEAM GENERATOR/BOILER CALIBRATION PROCEDUREFIGURE 9.4 - SOLENOID VALVE40 419.0 – REFERENCE BLOCKS (continued)Applies to STATIM 2000 and
FIGURE 9.6 - PUMPFIGURE 9.5 - CASSETTE42 439.0 – REFERENCE BLOCKS (continued)CASSETTE DIAGNOSTICS:If water is leaking at the drain from the cassette
FIGURE 9.7 - AIR COMPRESSORThe OEM recommends replacing the compressor filters every 6 months or every 500 cycles (whichever comes first).Flow44 459.0
46 472143658710 912 1114 13151719P1ThermocoupleMicroswitchSolenoidValveAir CompressorSteam Generator/BoilerInvensysWaterQualitySensorThermocoupleHeate
48 49FIGURE 9.9 - PLUMBING DIAGRAM9.0 – REFERENCE BLOCKS (continued)Female Quick ConnectFitting (Drain Line)ReservoirPumpIn-LineWater FilterSteam Gen
50 51ADC058 No OEM Part # Available CORD CLIP - 10/PKG • •CAF011 No OEM Part # Available RUBBER FOOT - 4/PKG • •RPB365 53-100518A 90º LIQUID TIGHT STR
52 53NOTES11.0 - RPI TOOLS & ACCESSORIES LIST(FOR PARTS TO FIT STATIM 2000 & 5000)RPA369 No OEM Part # Available THREADLOCKER 545 • •RPB792 No
Parts mentioned in this booklet are manufactured by Replacement Parts Industries, Inc. to fit SciCan STATIM® Cassette Autoclave Models 2000 and 5000.
23THISUPSIDEUPSIDETHISTHISUPSTOPSTARTLEVEL290FILL LINEMINIMUMGOLRKNNINCUOI2.1 – Extension Test CableRPI Part #SCT026(see page 4)2.2 – Start SwitchRPI
4 52.1 - EXTENSION TEST CABLE(RPI PART #SCT026)When testing, troubleshooting or repairing the STATIM au-toclave, install the RPI Extension Test Cable.
6 72.3 - PUMP TESTER BOTTLE(RPI PART #SCK024)The RPI bottle has a “swivel fitting” on the lid that allowsyou to control the position and balance of th
8 92.4 - SPANNER NUT(RPI PART #RPT364)When removing the solenoid flute/bonnet, it is suggestedby the OEM to use a pair of pliers on the welded area of
10 112.6 - CASSETTE SEAL REMOVALTOOL (RPI PART #RPT372)The Cassette Seal Removal Tool was designed to help remove the cassette seal for either mainten
12 13FIGURE 2.10BTIP - If realigning existingprobe bracket, loosen (DO NOTREMOVE) 4 mounting screwsand compression fitting con-nections to allow repos
TDS114 152.11 – WATER QUALITY TESTER(RPI PART #RPT820)The Water Quality Tester (RPI Part #RPT820) is a meter de-signed to measure water quality. This