3. Installation
3.10 Preparing Unit for Use
Once the unit is installed and before any instruments are sterilized, run two Wrapped Cycles
(see Section 4.8 Statim 2000 - Running a Cycle and 4.10 Statim 5000 - Running a Cycle).
Remove the cassette once it has cooled. Clean the top (lid) and bottom (tray) sections using
a soft cloth to wipe the inside surfaces and then rinse thoroughly with tap water. Once the
cassette is clean and dry, coat the inside surfaces with STAT-DRI.
3.11 Shipping the Unit
Before you move the unit, you will need to drain the reservoir. To do so, follow these steps:
1. Place a water container below the unit.
2. Using the drain tube (see Section 3.5 Priming the Statim Pump, Figure 6) empty the
contents of the reservoir into the water container.
3. Remove any remaining water from the reservoir with a non-linting, absorbant towel.
4. Screw-in the three leveler feet found underneath the unit.
5. Repack the unit in the original packing materials and include all accessories originally
included with the unit.
6. Specify heated and insured shipping.