4. Instructions For Use
4.8 Statim 2000 — Running a Cycle
To operate each cycle, follow these steps and watch the LCD .
or
or
or
14:23 3/11/2001
SELECT A CYCLE
SELECT A CYCLE
PRESS START
WARMING UP
UNWRAPPED
134˚C for 3.5 min.
WRAPPED
134˚C for 10 min.
RUBBER AND PLASTICS
121˚C for 15 min.
Turn the power switch
at the back of the unit to ON.
The LCD Display reads:
(if optional printer is installed)
Press the appropriate cycle
button on the keypad.
The display will read either:
As the button is released
the display reads:
The amber indicator light comes on to indicate that the cycle is in
progress. During the cycle, the phases of the cycle will be displayed
and the messages will appear similar to:
[
CYCLE YOU SELECTED
]
[
CYCLE YOU SELECTED
]
95˚C
130˚C
134˚C 212kPa 3:30
105˚C
CONDITIONING
PRESSURIZING
STERILIZING
VENTING
20
Comentários a estes Manuais