3. Installation
To fill the reservoir, follow these steps
(see Figure 5):
1. Remove the reservoir cap
n
2
2. Pour steam-process distilled water into the
reservoir until almost full (a maximum of
4L/1 U.S. gal). Use a funnel to avoid
spillage.
3. Replace and secure the cap.
3.5 Priming the Statim Pump
To prime the Statim pump, follow these steps (see Figure 6):
1. Move the unit to the edge of the work surface. The front leveler feet should be
approximately 12 mm/0.5˝ from the edge.
2. Lift the front left corner of the unit upward and remove the drain tube
n
3
from
the clip located on the underside of the unit.
3. Pull the drain tube outward so the free end can be positioned over a water container.
4. Fill the reservoir with steam-process distilled water.
5. Remove the plug
n
4
from the end of the drain tube and allow water to drain from the
tube into a container for 30 seconds. When the
water flows in a steady stream, replace the stopper.
6. Lift the front left corner of the unit upward and
reinsert the tube into the clip on the underside of
the unit. Push the excess length of tubing back
into the space provided.
Make sure the plug in the drain tube is secured.
After installation, and before sterilizing any instruments, run two Wrapped cycles.
For further instructions, see Section 3.9 Preparing the Unit for Use.
3.6 Setting the Time and Date
To set the time and date, follow these steps and watch the blinking cursor on the LCD:
1. Power the Statim OFF.
2. Press and hold the UNWRAPPED cycle button.
3. While pressing the UNWRAPPED cycle button,
power the Statim ON. The LCD display reads:
4. Use the cycle buttons to select and change the selected fields value. To increase a field’s
value, press the UNWRAPPED cycle button. Hold the button down to increase the value.
5. To decrease the value, press the WRAPPED cycle button.
n
2
Figure 5
Figure 6
n
3
n
4
14:23 11/15/2006
HH:MM DD/MM/YYYY