4. Instructions For Use
The sterilizing cycle ends when the reminder tone sounds, the amber light flashes and
the message CYCLE COMPLETE is displayed. The display appears similar to:
AIR DRYING 60:00
CYCLE COMPLETE
You may interrupt the automatic Air Drying phase at any time, or let it run to the end.
To ensure that Wrapped Instrument loads are dry, allow this cycle to run to completion.
If the cassette contains wrapped instruments and the wraps are not dry when
the cassette is opened, the instruments must be handled in an aseptic manner
for immediate use, or resterilized.
To remove the cassette, you must first press the button on the keypad.
The display appears similar to:
PLEASE WAIT
CYCLE COMPLETE
Be careful. The metal parts will be hot, and the cassette will contain hot steam.
After the cassette is removed from the unit, it should be opened to hasten drying
of unwrapped instruments.
If the Air Drying phase is not interrupted, the program will automatically continue to
run. When this phase ends, the display reads:
REMOVE CASSETTE
CYCLE COMPLETE
When a sterilization cycle is successful, the reminder tone sounds and the amber light
flashes until the STOP button is pressed or the cassette is removed from the unit.
If a message is displayed which gives a CYCLE FAULT code or a NOT STERILE
message, the cassette contents may not be sterile. See Section 8 Troubleshooting for
more information.