Message MSD NOT CONNECTED <>
INSERT MSD/FLASH
Message Missing lines of
Data on MSD/FLASH
Refer to the STATIM screen to confirm successful
sterilization.
Reset the Data Logger by unplugging its power,
disconnecting the MSD and waiting 10 seconds. Then
re-connect the power adapter and insert the MSD into
the Data Logger.
If the problem persists, contact the SciCan Service
Center.
Message: The file or directory in the MSD
is corrupted or unreadable.
Refer to the STATIM screen to confirm successful
sterilization.
The MSD may have been unplugged while data was
being written to it. The MSD should not be unplugged
until after “SAFELY REMOVE MSD <> MSD/FLASH
DETECTED” is displayed.
The corrupted files or directories may be lost. Reformat
the MSD on your computer.
Message MSD/FLASH FULL <>
REPLACE MSD
The MSD is full. Export the data.
Check the serial cable connection.
Check the power connection.
Ensure the lower red LED is lit.
Check that the Mass Storage Device is properly
inserted.
Repeat the instructions for Installing the SciCan Data
Logger on your Statim.