As the cycle is running, an intermittent buzzing sound will be apparent as the pump
injects water into the steam generator. A random clicking sound will also be heard as the
exhaust valve opens and closes.
Once the sterilization cycle is complete, a reminder tone will sound and the amber light
will flash as the cycle moves into the Air Drying phase.
The buzzing noise during the air drying stage is the compressor operating. The air
drying phase of the cycle may be interrupted at any time by pressing the STOP button.
The display will appear similar to:
[
CYCLE YOU SELECTED
]
REMOVE CASSETTE
When the automatic 60-minute air drying stage is finished the display reads:
[
CYCLE YOU SELECTED
]
CYCLE COMPLETE
If a sterilization cycle is successful, the reminder tone sounds and the amber light
flashes until the STOP button is pressed, or the cassette is removed from the unit.
Be careful. The metal parts will be hot, and the cassette will contain hot steam. After the
cassette is removed from the unit, it should be opened to hasten drying of unwrapped
instruments.
If a message is displayed which gives a CYCLE FAULT code or a NOT STERILE
message, the cassette contents are not sterile. See Section 8 Troubleshooting for
more information.