5. Maintenance
The schedules below describe the necessary actions.
Daily
Weekly
Every
6 months
Operator
Once a year
Technician
5.10 Preventative Maintenance Schedule
To ensure trouble-free performance, both the operator and the dealer must follow
a preventative maintenance schedule. NOTE: Please refer to your National, Regional,
State or Safety laws for any additional reoccurring user testing that may be required.
31
Water Reservoir Replace water as needed.
For opthalmic use, drain at the end of every workday, leave empty, and refill at the start
of the next workday.
Waste Bottle Empty the waste bottle every time you refill the water reservoir.
Fill the bottle with water, up to the MIN line marking. You may also add some
chlorine-free disinfectant.
Cassette Wash the interior of the cassette with dishwashing soap or a mild detergent that does not
contain chlorine.
Scrub the inside with a cleaning pad designed for use with Teflon
-coated surfaces.
After removing all traces of the detergent, treat interior surfaces of the cassette with
the STAT-DRI™ Plus drying agent to enhance the drying process. Order more STAT-DRI™
Plus from your dealer quoting 2OZPLUS, 8OZPLUST, or 32OZPL US.
Biological and/or Check the filter for dirt and moisture. Replace if dirty. Call for service if wet.
Air Filter
Water Filter Check the water reservoir filter every week and clean if necessary. Replace only if necessary.
Cassette Seal Replace every 500 cycles or six months (whichever is first), or whenever necessary.
Biological and/or Replace every 500 cycles or six months (whichever is first).
Air Filter
Cassette Check the tray, lid and seal for damage. Replace if necessary.
Biological Filter Inspect the biological filter for moisture.
Solenoid Valve Inspect the valve and clean if dirty. Replace the plunger if defective.
Pump Clean the filters, replace if dirty.
Check Valve Remove the exhaust tube from the back of the unit during the air drying phase. Check for air
coming from the fitting.
Remove the air compressor tube from the check valve inlet while running a cycle. Make sure
no steam is leaking from the valve. Replace if there are any leaks.
Water Reservoir Check the reservoir for dirt. Clean and rinse with steam process distilled water if necessary.
Calibration Calibrate the unit.