2.1 Disclaimers
Use only steam-process distilled water in your Statim. Deionized, demineralized, or
specially filtered water should not be used. Never use tap water.
Do not permit any person other than certified personnel to supply parts for, service or
maintain your Statim. SciCan shall not be liable for incidental, special or consequential
damages caused by any maintenance or services performed on the Statim by a third
party, or for the use of equipment or parts manufactured by a third party, including lost
profits, any commercial loss, economic loss, or loss arising from personal injury.
Never remove the cover of the unit and never insert objects through holes or openings in
the cabinetry. Doing so may damage the unit and / or pose a hazard to the operator.
All elements of this book are common to both Statim 2000 and Statim 5000 except
when noted.
IMPORTANT:
Follow local guidelines governing verification of the sterilization procedure.
Drying Performance
The Statim 2000 and 5000 have been designed to provide a complete sterilization
solution for your unwrapped and wrapped instrument needs: rapid sterilization
balanced with rapid drying, by using the SciCan Dri-Tec Drying Technology.
To dry instruments, the Statim 2000 uses a combination of forced filtered air and
convection heat. The convection heat is derived from utilizing the remaining heat in the
system after the sterilization phase has been completed. The heat is then captured and
released into the cassette to rapidly dry a properly loaded cassette.
The Statim 5000 also uses forced filtered air and heat generated by the sterilization
phase to dry the instruments. The heat generated by the sterilization phase is absorbed
by the drying plates and then transferred directly to the load. This then results in an
accelerated, rapid drying of a properly loaded Statim cassette.
Please refer to this operator's manual for instructions on the proper arrangement of
instruments in the cassette and the use of Stat-Dri plates (Statim 5000). By carefully
following these directions on how to properly load the cassette chamber, rapid drying of
the load will be achieved.